AirSep Rev K
AirSep Rev K
Used on: AirSep ???
Notes:
The most frequently found fault is damage to the two 1 Ohm, 2 Watt resistors located to the right of the Blue MOV in the left-hand picture (wrapped in heat shrink tubing in this picture).
No pressure sensor
AirSep CB084
AirSep CB084
Used on: AirSep ???
Notes:
The most frequently found fault is damage to the two 1 Ohm, 2 Watt resistors located to the right of the Red MOV in the left-hand picture.
AirSep CB100-1
AirSep CB100-1
Used on: AirSep ???
Notes:
???
AirSep CB153-1 Rev A
AirSep CB153-1 Rev A
Used on: AirSep Elite
Notes: